RP-HPLC Method Development and Validation for Assay of Levetiracetam in Tablet Dosage Form

 

J Valarmathy1*, L Samueljoshua2, G Rathinavel1, C Selvin Thanuja1 and T Sivakumar1

1Department of Pharmacy, Nandha College of Pharmacy, Erode 638052, Tamilnadu, India

2Department of Pharmacy, Padmavathi College of Pharmacy, Dharmapuri, Tamilnadu, India

 

*Corresponding Author E-mail:  valarmathyjoshua@rediffmail.com

 

ABSTRACT

A simple, reproducible and efficient reverse phase high performance liquid chromatography (RP-HPLC) method has been developed for estimation of Levetiracetam in its tablet dosage form. Separation was done by using mobile phase consists of buffer solution (pH 2.8) and acetonitrile in the ratio of 90:10. Chromatography separations were carried out on prontosil C18 column (150X4.6mn; 5µm) at a flow rate of 1.2 ml/min and UV detection at 215nm and the retention time for levetiracetam is 4minutes. The linear dynamic response was found to be in the concentration of 45µg-270µg/ml. The slope, intercept and correlation coefficient were found to be 0.9999 respectively. The percentage recovery of Levetiracetam was found to be 99.08%. Proposed methods were found to be simple, accurate, precise and rapid and could be used for routine analysis. This condition is applied only for tablet dosage form. The statistical parameters and recovery studies were carried out and reported.

 

KEY WORDS                 :  Levetiracetam, Estimation, Tablets, RP-HPLC.

 


 

INTRODUCTION:

Levetiracetam (Fig-1) is a single enantiomer of (­)-(s)- ethyl-2-oxo-1-pyrrolidine acetamide. Levetiracetam can prevent myoclonic jerks and generalized epileptiform activity in patients with photosensitive epilepsy. A few RP-HPLC methods have been reported in the literature for the estimation of levetiracetam. The authors now proposed a high degree of accuracy and precise HPLC method for the estimation in pure as well as tablet dosage forms.    

 

EXPERIMENTAL:

A prominence Shimadzu high pressure liquid chromatographic instrument provided with prontosil C18 column (150X4.6mm, 5µm) 10µl syringes and supported by window software was employed in the study. Levetiracetam was obtained from Orchid Chemicals Pharmaceuticals Ltd; Chennai. Potassium dihydrogen phosphates,                Sodium-1-heptane sulphonate, Acetonitrile and Water were of HPLC grade. All other chemicals were of Analytical Reagent grade.

 

Chromatographic conditions:

 The mobile phase was pH 2.8 buffer solution and acetonitrile in the ratio of 90:10(v/v). Mobile phase was filtered through 0.45µm membrane filter and sonicated before use. The mobile phase was pumped from the solvent reservoir in the ratio of 90:10(v/v) to the column at a flow rate of 1.2ml/min. The run time was set at 8 minutes. The column was maintained at ambient temperature and the volume of each injection was 10µl. Samples were injected in to the column and the absorptivity of the mobile phase was monitored at 215nm using a variable wavelength detector.

 

Assay procedure:

Preparation of pH 2.8 Buffer solution

Potassiumdihydrogen phosphate (0.01mol, 1.36gm) and Sodium-1- heptane sulphonate (0,61gm) in 1000ml of milli equivalent water was solubilised and adjusted the pH to 2.8 ±0.05 with ortho phosphoric acid solution.

 

Preparation of diluents solution:

Mobile phase consisting of buffer solution (pH 2.8) and acetonitrile in the ratio of 90:10 was used as diluents and were filtered through 0.45µ membrane filter and sonicated for 10 minutes.

 

Preparation of standard solution:

About 90mg of pure sample of levetiracetam was weighed accurately and transferred to a 50ml volumetric flask and dissolved in 35ml of diluent. The solution was sonicated for 30min and then the volume was made up with a further quantity of diluent to get 1.8mg/ml solution. From this 5ml of the solution was transferred in to 50ml volumetric flask and the volume was made up to mark with diluent to get 180µg/ml. Subsequent dilution of this solution ranging from 45-270µg/ml were made in 10ml volumetric flask. The prepared solutions were filtered through 0.45µ membrane filter and then 10µl of filtrate was injected each time in to the column at a flow rate of 1.2ml/min. Each concentration was injected 6 times into the column and corresponding chromatograms were obtained. Detection of the drug was one at 215nm from the chromatogram, the retention time and mean peak area rations were recorded for all the concentration. These determine the linearity by applying the mentioned chromatographic conditions. From the data obtained, correlation coefficient, Y-intercept and slope were calculated to provide mathematical estimation of the degree of linearity. The results are shown in table-1

 

 

Table1- Analytical performance parameters of Levetiracetam

                Parameters

              Average area

Linear dynamic range µg/ml

Correlation coefficient

Y-intercept

Slope of regression line

              45-270µg/ml

              0.9999

               12438

               6474.3x

 

Estimation of Levetiracetam in tablet dosage forms:

20 tablets were weighed and powdered. Accurately weighed portion of the tablet powder equivalent to 180mg was taken in 100ml volumetric flask and 75 ml of diluent was added, shaken well and allowed to stand for 30min with intermittent sonication to ensure complete solubility of the drug. The mixture was then thoroughly mixed and made up to the mark with diluent. Centrifuged at 2500rpm for 10min. From the above supernatant solution, pipetted out 5ml into a

 

50ml volumetric flask, diluted to volume with diluent, mixed well and filtered through a 0.45µ membrane. 10µl of filtrate was injected into the column. The concentration was injected 6 times. The amount of drug present in the tablet formulation was calculated by comparing the peak area from the standard.

 

The amount of Levetriacetam present in the sample =

 

Sample peak area    standard dilution     P

--------------------- × ---------------------× --× Average weight of tablet

Standard peak area  Test dilution         100

 

P = % potency of Levetiracetam working standard

 

The HPLC method developed in the present study has also been used to quantify Levetiracetam in tablet dosage form. Levetiracetam tablets (containing 750mg) were quantified using the proposed analytical method. No interfering peaks were found in the chromatogram indicating that the tablet excepients did not interfere with the estimation of drug by the proposed HPLC method. The tablets were found to contain 97.00-103.00% of drug. A known amount of drug solution was added to the sample of tablet dosage form and subjected to estimation of drug by proposed method. There was a high recovery of (Table 2) levetiracetam (97.15-101.41%) indicating that the proposed procedure for determination of levetiracetam in tablet dosage form is highly accurate.

 

RESULTS AND DISCUSSION:

The aim of this study was to develop a simple, accurate and precise HPLC method for the analysis of levetiracetam in bulk and tablet dosage forms using mobile phase (pH 2.8 buffer and acetonitrile 90:10 v/v) and commonly employed                     RP- C18 column with UV detector at 215nm.

 

Results of recovery study -Table 2

 

Spike level

%

 

 

Amount of  Levetiracetam

 

Recovery from drug   solution

Recovery from formulation

 

Mean amount

 

Mean (%) recovery

 

Mean amount

 

Mean(%) recovery

25

50

75

100

150

187.5

375

562.5

750

1125

  189.5

  378.6

  567.01

  755.6

  1133.88

  98.34

  97.09

  97.51

  99.13

  97.70

  185.52

  368.21

  576.85

  745.63

  1120.57

  97.22

  97.15

  101.41

   98.42

   97.06

 

 

Table 3 -Peak Area

Name

                Peak area

Standard Levetiracetam                                   Sample   Levetiracetam

                1173908

                1172850

 

 

The run time of the method was set at 8mintues. Retention time was found to be around 4minutes and levetiracetam appeared on chromatogram at 4.170min. This indicates that the present HPLC method is rapid which in turn shows that the method consumes less volume of HPLC solvents. When the same drug solution was injected 6 times, the retention time of the drug was found to be same. The retention time of the standard levetiracetam was observed to be 4.174minutes. The quantitative estimation was carried out on tablet by taking the same concentration as for standard solution and corresponding sample peak area were represented in Table 3. The average percentage purity was found to be 99.08%Hence, the proposed HPLC method is simple, precise, accurate and rapid for the determination of levetiracetam in dosage forms. It can be easily and conveniently adopted for routine quality control analysis of the drug.

 

ACKNOWLEDGEMENTS:

The authors thank Orchid Chemicals and Pharmaceuticals Ltd; Chennai, India for providing a gift sample of levetiracetam. The authors are also thankful to Mr. V. Shanmugam, Chairman, Nandha College of pharmacy, Erode-52, Tamilnadu, India for providing facilities for the research work

 

REFERENCES:

1.        Snyder, L.R., Practical HPLC and method development, 2nd Ed.; 1997,175.

2.        Jeffery’s, Bassett.J, and Mendham.J., Vogel’s text book of quantitative chemical analysis, 5 th Ed.; 1994,219.

3.        Frank settle, Handbook of instrument techniques for analytical chemistry, 2004,151.

4.        Harold.F.W. and Jorge Reyes, Modern chemical analysis and instrumentation 2001,256.

5.        Skooge, Holler; and Crouch, Fundamentals of analytical chemistry, 8 th Ed,; 2004,976.

6.        Willard merit, H.H., Dean, Jr., and Dean, J.A., Instrumental method of analysis 6 th Ed.; 1986,504.

7.        Satinder.A and Stephen scypinski, Handbook of modern pharmaceutical analysis, 2001, 3,419.

8.        Physician desk reference, 60 th Ed.; 2006, 3307.

9.        www.Rxlist.com/cgi/generic3/keppra.htm

10.     Jens martens –Lobenhoffer.J and Bode-Boger.S.M., J.chromatogr B Analyt Technol biomed Life sci.2005, 819(1), 197.

11.     Rao.B.M. Ravi.R., Shyam Sunder Reddy, B.Siva kumar, S.Gopichand, J pharm Biomed Anal.200435 (5), 1017.

12.     Tiegong guo, Lisa Oswald, M.Damodara Rao Mendu and Steven solidin.J, Clin chim Acta.2007, Jan 375(1-2), 115.

13.     Isoherranen.N, Roeder.M, Soback.S.Yagen.B, Schurig.V and Bialer.M, J chromatogr B Biomed sci Appl.2000, Aug 18,745(2), 325.

14.     Rathnaraj.N and Doheny.HC, the Drug Monit, 1996, Apr, 18(2), 154.

15.     ICH Harmonized Tripartite guidelines, validation of analytical procedure, methodology, 1996, Nov 6 th, 3.

16.     ICH Harmonized Tripartite guidelines, validation of analytical procedure, methodology, 1994, Oct, 5.

17.     United States pharmacopoeia validation of compendial

Method, 28thEd; 2005, 2440.

18.     British Pharmacopoeia, validation procedure, 2005, 4, A-456.

 

 

 

 

 

Received on 29.07.2008       Modified on 06.08.2008

Accepted on 14.10.2008      © RJPT All right reserved